Dismantling the Compact Breathing System
Dismantling the Compact Breathing System
Remove all breathing hoses
Remove the ventilation hose.
Remove the fresh gas hose from the breathing system.
Remove the scavenger hose.
Remove the flow sensor cable.
Remove the O2 sensor cable.
Remove the breathing pressure cable.
Remove the COSY shielding, being careful to guide any cables from the slots in the side of the cover.
Detach the APL-bypass and the PEEP/PMAX lines from the breathing system and from the side of the machine.
Remove the absorber.
Remove the compact breathing system.
Remove plug for the inspiration dome.
Dismantling the Inspiratory and expiratory Valves
Unscrew the retaining nuts.
Remove the inspection caps.
Extract the valve discs.
Remove the gaskets on top of the valve discs.
Removing the Flow Sensor
Loosen fitting on the expiration port.
Remove the flow-sensor guard.
Extract the flow sensor.
Dismantling the APL Valve
Unscrew the retaining nut.
Remove the APL valve.
Unscrew the waste gas outlet port.
Dismantling Parts of the Ventilator
Open the ventilator door.
Disconnect the ventilator chamber pressure sensor line from the ventilator chamber pressure sensor line port.
Unlock the three clasps.
Remove the cover.
Remove the ventilator diaphragm.