Clean and disinfect the machine and components before each service (and also when returning for repair).
When Required
Replace the O2 sensor when calibration is no longer possible.
Replace the flow sensor when calibration is no longer possible.
Replace the pressure measuring line (silicone rubber hose and sleeve).
Replace APL-bypass and PEEP and PMAX silicone rubber hoses.
Replace the ventilator hose if it is discolored or damaged.
Replace the AGS filter when blocked
Every 12 Months
Inspection and service by trained service personnel. Dräger Medical recommends DrägerService.
Fabius GS premium
Breathing systems
Vaporizers
Sensors
Ventilator hose
Replace the bacterial filter on the pressure measuring line.
Replace the diaphragm in the ventilator (patient).
Replace vaporizer O-Rings
After 2 Years
By trained service personnel:
Replace diaphragm and sealing ring of PEEP valve.
Replace diaphragm and sealing ring of fresh gas decoupling valve.
Replace diaphragm and sealing ring of APL bypass valve.
Replace SpiroLife flow sensor.
After 3 Years
By trained service personnel:
Replace the lead gel rechargeable battery for the back-up power supply.
Replace the diaphragm and O-rings of the ventilator (piston).
Replace seals of breathing system canister assembly.
After 6 Years
By trained service personnel:
Replace pressure reducer.
NOTE:
If desired, the customer may receive a list of those parts and their assembly instructions to be replaced in the event of repairs being necessary.