Clean and disinfect the machine and components before each service (and also when returning for repair).

When Required

  • Replace the O2 sensor when calibration is no longer possible.
  • Replace the flow sensor when calibration is no longer possible.
  • Replace the pressure measuring line (silicone rubber hose and sleeve).
  • Replace APL-bypass and PEEP and PMAX silicone rubber hoses.
  • Replace the ventilator hose if it is discolored or damaged.
  • Replace the AGS filter when blocked
  • Every 12 Months

    Inspection and service by trained service personnel. Dräger Medical recommends DrägerService.

  • Fabius GS premium
  • Breathing systems
  • Vaporizers
  • Sensors
  • Ventilator hose
  • Replace the bacterial filter on the pressure measuring line.
  • Replace the diaphragm in the ventilator (patient).
  • Replace vaporizer O-Rings
  • After 2 Years

    By trained service personnel:

  • Replace diaphragm and sealing ring of PEEP valve.
  • Replace diaphragm and sealing ring of fresh gas decoupling valve.
  • Replace diaphragm and sealing ring of APL bypass valve.
  • Replace SpiroLife flow sensor.
  • After 3 Years

    By trained service personnel:

  • Replace the lead gel rechargeable battery for the back-up power supply.
  • Replace the diaphragm and O-rings of the ventilator (piston).
  • Replace seals of breathing system canister assembly.
  • After 6 Years

    By trained service personnel:

  • Replace pressure reducer.
  • NOTE:

    If desired, the customer may receive a list of those parts and their assembly instructions to be replaced in the event of repairs being necessary.