Controlled ventilation
In controlled ventilation, the TurboVent2 rotates generating a small circular flow (0.6 LPM in expiration, 1.2 LPM in inspiration). The circle flow enables free spontaneous breathing even on high pressure levels and quick gas mixing in the breathing system. The APL valve is bypassed by the APL bypass valve. The Pmax/PEEP valve is active and synchronised with ventilator breath rate. It controls the inspiratory pressure during inspiration and PEEP during expiration. The blower rotates depending on the set values (Pinsp or VT) in PC or VC. In volume controlled, the TurboVent2 increases its speed till the set volume, measured by the flow sensor, is reached in inspiration. In pressure control, TurboVent2 increases its speed as well till the desired set inspiratory pressure achieved. In all ventilation modes, the manual bag serves as a reservoir for the re-breathed gas. During inspiration, the blower uses this gas including the fresh gas.