Disassembly of the breathing system components
Sequence of disassembly
Remove reusable CO2 absorber:
- Remove the CO2 absorber from the breathing system.
- Remove and dispose of the soda lime dust filter.
- Empty the CO2 absorber.
- Remove the absorber insert from the absorber container. Leave the inner and outer sealing rings on the absorber insert.
Reassembling of components is carried out in a reverse order.
Information concerning dismounted accessory parts and attached devices
Consult the appropriate instructions for use for the article concerned:
Disposal of accessories
When disposing of the following accessory parts, observe the hospital hygiene regulations and the respective instructions for use:
- flow sensors
- breathing hoses
- filter, HME, HMEF
- manual breathing bag
- masks
- water trap
- CLIC absorber, Infinity ID CLIC absorber
- soda lime
Dispose on the following articles according to hospital hygiene regulations:
- sample line
- soda lime dust filter
- AGS filter